In vitro fertilisation Incubator MEMMERT IVF module for ICO50med

csm_ICO50med_IVF-Modul

Standard delivery IVF module for ICOmed

Deskripsi

SPECIFICATIONS

Temperature

  • Working-temperature range 5 °C above ambient temperature up to +50 °C
  • Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
  • Setting temperature range +18 to +50 °C
  • Setting accuracy temperature 0.1 °C
  • Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
  • Temperature variation in chamber at + 37 °C (to DIN 12880:2007-05) +/- 0.3 K
  • Temperature fluctuations with time at 37 °C (to DIN 12880:2007-05) +/- 0.1 K
  • Humidity

  • Humidity control (standard) Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
  • Setting accuracy humidity 0.5 % rh
  • Setting range active humidity control (with option K7) 40 to 97 % rh and rh-Off
  • Control of standard components

  • CO2 control Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
  • Adjustment range CO2 0 to 20 % CO2
  • Setting accuracy CO2 0,1%
  • Variation in time CO2 +/- 0.2 % CO2
  • Adjustment range O2 (with option T6) 1 to 20 % O2
  • Setting accuracy O2 0.1 % O2
  • Stainless steel interior

  • Dimensions w(A) x h(B) x d(C): 400 x 425 x 330 mm (d less 35 mm for fan)
  • Interior material 1.4301 (ASTM 304), corrosion resistant
    Volume 56 l
  • Max. number of internals 5
  • Max. loading of chamber 75 kg
  • Max. loading per internal 15 kg
  • In vitro fertilisation Incubator MEMMERT IVF module for ICO50med
    In vitro fertilisation Incubator MEMMERT IVF module for ICO50med